Compact Thermologger AM-8101E / 8101K

コンパクトサーモロガー AM-8000E / 8000K

Measurement Range
E -200 to 800°C / K -200 to 1370°C

Number of inputs
6ch

Interface
RS-232C

Input Type
Thermocouple Input : Type E, K (typeJ,T,R(option))

Interval
Standard model : 1 second to 99 minutes 59 seconds
High-speed mode (6ch.model) : 0.10 seconds to 0.99 seconds

(Temperature probes not included)

Download catalog

Specifications

ModelAM-8101
Channel6ch : ANP connector (Miniature Plug)
Input TypeThermocouple Input: Type E, K (typeJ,T,R(option))
Measurement Range(Table 1)
Resolution0.1°C/ 1°C (fixed)
Measurement Accuracy(Table 2)
Memory capacity60,000 data
High-speed interval0.1 second
InterfaceRC-232C
Dimensions / Weight92(W) × 155(H) × 36(D) mm [Protruded parts excluded] / approx.380g [Including dry batteries]

Reference junction compensation accuracy

±0.3°C at 25°C ± 10°C
Temperature coefficient±(0.01% / °C) of measurement range except 25°C ± 10°C
Time interval

Standard model : 1 second to 99 minutes 59 seconds

High-speed mode (6ch.model) : 0.10 seconds to 0.99 seconds
Signal source resistance

Temperature

standard mode : 500Ω or less

high-speed mode : 100Ω or less

Safe excitation voltage

Between input pins on same channels : ±10V DC (temperature), ±30V DC (voltage)

Between input pins on adjacent channels : ±60V AC

LlinearlizerDigtal Linearlizer method (compliant with IEC 60584-1 (2013))
DisplayGraphic liquid crystal display, 128×64 dots (55×27mm)
Environment operational/Environment storage

0 to 40°C, 20 to 80%RH (Without Dew)

-10 to 50°C, 10 to 90%RH (Without Dew)

Power supply4 AA Alkaline batteries or AC adapter (Option accessories)
Supplied accessories4 AA Alkaline batteries, Test report, Manual, Softcase, Communications cable (RS-232C)
Compatible standardsCE Marking, RoHS


Table 1 : Measurement Range

Type1°C Range0.1°C Range
E-200 to 800°C-100.0 to 500.0°C
K-200 to 1370°C
J-200 to 1100°C
T-200 to 400°C-100.0 to 400.0°C
R0 to 1760°C-

*Switches automatically to 1℃ resolution range when measured temperatures fall outside the 0.1℃ resolution range.

 

Table 2 : Measurement accuracy(Not including reference junction compensation accuracy, at 25°C ± 10°C)

TypeResolutionMeasurement rangeMeasurement accuracy
Type E, K, J, T1°C0°C to± (0.1% of reading + 1°C )
-100°C to -1°C± (0.5% of reading + 1°C )
-200°C to 101°C± (1% of reading + 2°C )
0.1°C0.0°C to± (0.1% of reading + 0.3°C )
-100.0°C to -0.1°C± 0.5 °C
Type R1°C0 to 1760°C± (0.1% of reading + 3°C )

 

6ch model Continuous use time

Interval0.1 sec1 sec10 sec10 min
Battery lifeapprox. 20happrox. 40happrox. 80happrox. 240h

* When operated with alkaline AA-size dry batteries at 25°C.
* For long-term measurement, please use optional AC adapter.
* Please contact us for details on available time.

Optional accessories

画像

Special data analysis software

AMS-850

画像

AC adapter

AD-115-500-HR-R (for AC115V)
AD-220-500-HR-R (for AC220V)

画像

Carrying case

AM-CC2

About repair

This product is available for repair. Please ask our dealer.