Temperature Calibration System ACSⅡ-4000

画像

Surface / Internal temperature calibration system

Surface Range:30 to 500°C
Internal Range:30 to 500°C

Download catalog

Overall Accuracy

ModelACSII-4000
Calibration ObjectSurfaceInternal
Range30.0 to 100.0°C+0.4°C±0.26°C
-0.6°C
100.1 to 200.0°C+0.3°C±0.32°C
-0.9°C
200.1 to 300.0°C+0.1°C±0.37°C
-1.1°C
300.1 to 400.0°C-0.2°C±0.43°C
-1.5°C
400.1 to 500.0°C-0.9°C±0.48°C
-2.2°C


Control Unit

ModelACSII-4000
Range 30 to 500℃
Control PrinciplePID auto-Tuning Control
ControlInputPt100Ω
Accuracy±0.2% of reading+0.3°C
Display8mm high LED display
Reference sensorInputPt100Ω Φ4.8mm JIS C 1604
Accuracy±(0.05% of reading+0.05)°C
Display12.5mm high LED display
Power supply *AC100V~240V
Power Consumption1600W
Dimensions(W×H×D)430*177*350mm
Weight5kg

* Please select at purchase
 

Calibration Unit

ModelACSII-4000
Point8 holes based on specified dimension
Surface±0.1℃
Heater1500W
Overheating prevention mechanismBimetal breaker
Heating time30 → 250°C 60min
30 → 500°C 150min
Cooling time500 → 250°C 220min
250 → 100°C 450min
Dimensions(W×H×D)450*405*400mm
Weight70kg

 

 

Supplied accessories

ModelACSII-4000
Supplied accessories

Power cable,Heater cable, Sensor cable, Manual,Test report,

Protective cover for calibration block,

Standard sensor for calibration(Pt 100Ω platinum resistor)



About repair