Specifications of MG series / Magnet Probes

MG-1*/MG-2*/MG-6*

Model number*1MG-11
MG-13
MG-21
MG-23
MG-22
MG-24
MG-62
MG-64
Thermocouple typeType E or K
Temp. range*2-50 to 250°C-50 to 250°C-50 to 400°C-50 to 400°C
Tolerance*3-50°C ≦ t ≦ 100°C±2.5°C±2.5°C±2.5°C±2.5°C
100°C < t ≦ 200°C±3.1°C±3.1°C±3.1°C±3.1°C
200°C < t ≦ 300°C±4.0°C(less than 250°C)±4.0°C(less than 250°C)±5.0°C±5.0°C
300°C < t ≦ 400°C--±7.2°C±7.2°C
Response time*42s
Durability*5More than 50,000 contacts
Magnetic attraction700g400g400g650g
Guard materialsTeflon--Glass cloth board with stainless steel cover
Body materialsPhenol ResinTeflonPolyimideSumika super
Repair*6Repairable

MG-3*

Model number*1MG-33MG-34
Thermocouple typeType E or K
Temp. range*2-50 to 250°C-50 to 400°C
Tolerance*3-50°C ≦ t ≦ 100°C±2.5°C±2.5°C
100°C < t ≦ 200°C±3.1°C±3.1°C
200°C < t ≦ 300°C±4.0°C(less than 250°C)±5.0°C
300°C < t ≦ 400°C±7.2°C
Response time*42s
Durability*5A(More than 50,000 contacts)
Magnetic attraction340g
Body materialStainless(SUS303)Stainless(SUS303)
Repair*6Repairable

*1 The asterisk (*) is replaced by the number of the model name you selected. The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is available at -50°C or above within the operating temperature on a stationary flat and smooth metal surface.
*4 The response time is the time required to detect 99% of the true value on a flat and smooth metal surface.
*5 Number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface at a temperature of 300°C (or at the operating temperature limit if the operating temperature limit is below 300°C)
*6 Please refer to page repair.