Specifications of AX Series / General Stationary Surface Probes(Contact support type)

AX-*1*/AX-*2*/AX-*3*/AX-*4*/AX-17*

Model number*1AX-*1*AX-*2*AX-*3*AX-*4*
Thermocouple typeType E or K
Temp. range*2-50 to 300°C-50 to 400°C-50 to 500°C-50 to 800°C
Tolerance*3-50°C±2.5°C±2.5°C±2.5°C±2.5°C
0°C±2.5°C±2.5°C±2.5°C±2.5°C
100°C±2.5°C±2.5°C±2.5°C±2.5°C
200°C±2.5°C±2.5°C±2.5°C±2.5°C
300°C±2.5°C±2.5°C±2.5°C±2.5°C
400°C±3.0°C±3.0°C±3.0°C
500°C±3.8°C±3.8°C
600°C±9.0°C
700°C±10.5°C
800°C±16.0°C
Tolerance calculation method
t:Temperature(°C)
At within temperature range
1)-50°C ≦ t < -40°C:±2.5°C
2)-40°C ≦ t ≦ 500°C:±2.5°C or ± (0.0075 × | t |)°C, whichever is greater
3)500°C < t ≦ 700°C:±(0.015×|t|)°C
4)700°C < t ≦ 800°C:±(0.02×|t|)°C
Response time*41.5s
Durability*5More than 75,000 contacts
Pipe materialStainless(SUS316)
Grip materialPolyacetal
Movable range of the head±4°
Repair*6Repairable

 

AX-*5*/AX-*6*

Model number*1AX-*5*AX-*6*
Thermocouple typeType E or K
Temp. range*2-50 to 200°C-50 to 300°C
Tolerance*3-50°C±2.5°C±2.5°C
0°C±2.5°C±2.5°C
100°C±2.5°C±2.5°C
200°C±4.0°C±4.0°C
300°C-±6.0°C
Tolerance calculation method
t:Temperature(°C)
At within temperature range

1)-50°C ≦ t < -40°C:±2.5°C
2)-40°C ≦ t ≦ 300°C:±2.5°C or ± (0.02 × | t |)°C, whichever is greater

Response time*43.5s
Durability*5More than 75,000 contacts
Movable range of the head±4°
Pipe materialStainless(SUS316)
Grip materialPolyacetal
Repair*6Repairable

*1 The asterisk (*) is replaced by the number of the model name you selected. The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is available at -50°C or above within the operating temperature on a stationary flat and smooth metal surface. For tolerance at temperature not listed in the above table, please refer to calculation method of tolerance.
*4 The response time is the time required to detect 99% of the true value on a flat and smooth metal surface.
*5 Number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface at a temperature of 300°C (or at the operating temperature limit if the operating temperature limit is below 300°C)
*6 Please refer to page repair.